Reflectance spectrophotometer TR 500

Reflectance spectrophotometer TR 500
Ergonomic and flexible reflection spectrophotometers for measuring samples in different industries including food, medicine, coatings and plastics. The on-board camera-based locator and the black and white calibration plates ensure precise measurements and results at all times.

  • CMOS sensor (256 picture elements, double array)
  • Intuitive 3.5 inch TFT color LCD (capacitive), symbol-based
  • Free PC-based OnShade software which allows graphs and statistical process control
  • Operation languages: English, French, German, Spanish, Portuguese, Chinese
  • Customer-specific apertures
  • External memory with 2000 standards and 20000 samples


Scope of supply: Reflectance spectrophotometer, OnShade PC software, black and white calibration board (white tile and black interface), dust cover, USB cable, external DC power supply, USB stick with user manual in the operating languages, carrying case
Specifications
Optical geometry:di: 8°, de: 8°
Specular component:SCI, SCE
Colour space:CIE Lab, XYZ, Yxy, LCh, CIE LUV, Hunter Lab
Colour difference formular:ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc (2:1), ΔE*cmc (1:1), ΔE*00v, ΔE (Hunter)
Other colorimetic index:WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), TI (ASTM, E313, CIE/ISO), metamerism index (MI), staining fastness, colour fastness and strength, opacity
Spectral range:400 ... 700 nm
Spectral interval:10 nm
Memory:2000 standards, 20000 samples
Interface:USB, Bluetooth® 4.0
Light source:combined LED light
Illuminant:D65, A, C, D50, D55, D75, F2, F7, F11
Repeatability:MAV/SCI: ΔE ≤0.05
Inter instrument agreement:MAV/SCI: ΔE ≤0.2
Dimensions (W x D x H):184 x 105 x 77 mm
TypeTR 500
Aperture mm8
PK1
Cat. No.11409481
-
TypeTR 500
Aperture mm4
PK1
Cat. No.11409480
-
TypeTR 500
Aperture mm1 x 3
PK1
Cat. No.11409479
-
Alternate products:
Reflectance spectrophotometer TRA 500
Reflectance spectrophotometer TR 520
Reflectance spectrophotometer TRA 520
Reflectance spectrophotometer TR 515