The pH/Redox/ISE portable meter is used to determine the pH, the redox potential and the temperature. It combines the advantages of mobile field work with the precision and the comfort of a laboratory meter. The PHT 830 has an automatic temperature compensation for the pH measuring as well as an adjustable reference temperature with measurements without temperature sensor. For calibration a manual or automatic two point calibration routine can be used.
PHT 830 Set 1 incl. plastic electrode AT 830 pH K and 1.50 m cable, suited for universel applications in laboratories and field. PHT 830 Set 2 incl. glass electrode AT 830 pH G and 1.00 m cable, suited for applications in laboratories. PHT 830 Set 3 incl. penetration electrode AT 830 pH E and 1.00 m cable, suited for applications in food industries.
All electrodes are equipped with an integrated temperature sensor Pt 1000, allowing parallel measurement of pH value and temperature.
Scope of supply: pH meter incl. electrode as indicated, 1 x buffer solutions pH 4.006 and pH 6.865 each, 1 x KCl solution, case Specifications | | Measuring range / Resolution | | pH: mV: Temperature: | 0 ... 14 pH / 0.01 pH -1999 ... 1999 mV / 1 mV -10 ... 100 °C / 0.1 °C | Accuracy | | pH: Temperatur: | ± 0.03 (± 2 pH-steps) ± 0.1 °C | Display: | 128 x 64 pixel, backlighted | Interface: | USB with "Interface HMG USB", isolated | Ambient temperature: | -10 °C ... 55 °C | Rel. humidity: | Max. <95 % (non-condensing) | Connections: | Connector BK for sensor (4-pole USB interface socket) | Housing: | ABS | Dimensions (W x D x H): | 200 x 40 x 95 mm | Batteries: | 3 batteries Type AA, IEC R6, LR6, 1.5 V | IP code: | IP 65 |
|